DC current calibration
After the 1M
Ω
resistance point has been calibrated, the unit will prompt you for 10mA. Fol-
low these steps for DC current calibration:
1.
Connect the calibrator to the AMPS and INPUT LO terminals of the Model 2000 as
shown in Figure 2-3.
2.
Calibrate each current step summarized in Table 2-5. For each step:
• Set the calibrator to the indicated DC current, and make sure the unit is in operate.
• Make sure the Model 2000 display indicates the correct calibration current.
• Press ENTER to complete each step.
• Allow the Model 2000 to finish each step.
NOTE
If you are performing DC-only calibrator, proceed to the “Setting calibration dates”
paragraph.
Table 2-5
DC current calibration summary
Calibration step
Calibrator
current
Allowable range
10mA
100mA
1A
10.00000mA
100.0000mA
1.00000A
9mA to 11mA
90mA to 110mA
0.9A to 1.1A
2000 MULTIMETER
!
R
Output HI
Input
LO
Output
LO
Model 2000
5700A Calibrator
Amps
Note: Be sure calibrator is set for
normal current output.
Calibration
2-11
Figure 2-3
Connections for
DC and AC amps
comprehensive
calibration