Table 1-6
Limits for resistance verification
Ω
Range
Nominal
applied
resistance
Nominal reading limits
(1 year, 18°C-28°C)
Recalculated limits
Low limit
High limit
100
Ω
1k
Ω
10k
Ω
100k
Ω
1M
Ω
10M
Ω
100M
Ω
100
Ω
1k
Ω
10k
Ω
100k
Ω
1M
Ω
10M
Ω
100M
Ω
99.9860 to 100.0140
Ω
0.999890 to 1.000110k
Ω
9.99890 to 10.00110k
Ω
99.9890 to 100.0110k
Ω
0.999890 to 1.000110M
Ω
9.99590 to 10.00410M
Ω
99.8470 to 100.1530M
Ω
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
2000 MULTIMETER
!
R
Model 2000
5700A Calibrator (Output 2-wire Resistance)
Output
HI
Input
LO
Output
LO
Note : Use shielded cables to minimize noise.
Disable calibrator external sense mode.
Sense HI
Sense LO
Input
HI
Performance Verification
1-13
Figure 1-6
Connections for
resistance verifi-
cation (100M
Ω
range)