Page 40
DOC0945 (22 August 2022 15:28)
www.keit.co.uk
Back to Table of Contents
APPENDIX 1: Controller Specifications
Physical details
Parameter /
Feature
Description
Dimensions
See drawings below
Optional mounting adaptors are available to fit a 2u 19” rack.
Operating
Environment
The controller must be operated in a safe, non-hazardous environment.
•
T
ambient
: 0°C to 50°C
•
Relative humidity: 5% to 90% at 25°C
Monitor
Resolution
The recommended monitor resolution is 1280 x 1045 pixels or higher.
Mains Power
90 to 264 Volts AC, 47 to 63Hz
Power consumption 20W (internal power supply rated to 60W)
Communications
Link
Data communications between controller and spectrometer are
through a multimode duplex fibre-optic lead with dual LC connectors
available from Keit.
The maximum possible length is determined by the fibre optic cable
type used:
Type OM1 (62.5/125μm):
max length 275m
Type OM3
(50/125μm):
max length 500m
Keit offers an optional OM3 extension with ASM0627-09-Z.