8. Slowly move the device backwards and allow the patient’s
eye to rest for a few seconds, maintaining the alignment
position.
9. When the patient is ready for another reading, move the
device closer until the alignment cross appears again and
another reading is triggered.
10. Repeat the previous steps for further readings until the
measurement averaged from the individual readings is
acceptable.
11. When two consecutive readings are within 1mmHg an
audible high pitch notification will be heard indicating that
sufficient readings may have been taken (if sounds are
enabled in the menu settings). If successive readings are not
within 1mmHg of each other, Keeler recommends taking up
to four readings and using the average.
12. Press the OD/OS button to switch from left to right eye, or
vice-versa.
13. Press and hold the OD/OS button to clear all the readings.
14. Pressing the print button will produce a paper printout from
the printer installed in the docking station. The infra-red
window in the docking station and hand unit should be
unobstructed, and aligned within 1m (3 feet). The Docking
Station LED will flicker during the transmission of the IR data
and then extinguish when printing.
15. If the docking station is connected to a computer, pressing
the print button will export the raw data to the computer, as
long as the serial port has been enabled as described below.