Disposable Probe Adapter 

Part Number 2508-P-8026

Instructions for use

 

The system automatically initiates a purging  

 

cycle of 90 seconds. During the purge cycle an  

 

animated Wait Symbol is displayed alongside

 

the Probe Symbol.

 

Three short beeps signal that the purge cycle is  

 

complete. The equipment is now ready for use  

 

as indicated by the timer and the Ready Symbol. 

Freezing cycles can now be controlled by the Cryomatic 
footswitch as normal (see Cryomatic user manual).

Warnings and Precautions

•   Only to be used in conjunction with Keeler  

 

  Cryomatic (2509-P-1000) and Disposable Cryo    
  Probe (2508-P-7022)
•   Check probe before use – do not use or repair
  if damaged or deformed
•   Disconnect probe from adapter only when    
  system is depressurised

•   Maximum pressure 5860kPa (850psi)
•  Consult Cryomatic instruction manual if in doubt
•   Unauthorised interference invalidates warranty
•   This device should be used by trained staff    
  experienced in the use of ophthalmic equipment
•   USA federal law restricts this device to sale or  
  on the order of a physician

1

Connect the Probe Adapter to the Cryomatic console 
by aligning the coupling on the console spigots 
and pushing to a positive click (It is impossible to 
connect the coupling the wrong way round – correct 
orientation is with the release button uppermost). 

Connect the disposable probe by inserting the probe 
plug into the adapter and pushing the probe firmly
and squarely against the collar(1). 

The safety latch(2) will operate and the locking
ring(3) will spring towards the probe to confirm that 
the probe has been successfully connected. 

  

Operation

2

Before use ensure that the rear safety 
clip(4) is in the fully down position 
- under no circumstances should it 
be used if the safety clip is not in 
position (refer to image on page 2). 

1

2

4

3

 

Sterilized using Ethylene Oxide

Manufacturer

Caution

The CE mark on this product indicates
it has been tested to and conforms
with the provisions noted within the
93/42/EEC Medical Device Directive

Do Not Reuse

Use By

Batch Code/Lot Number/Batch Number

Consult Instructions For Use

This Symbol on the Product or on its 
Packaging and instructions indicates 
that it was put on the market place
after August 2005 and that this 
product shall not be treated as 
Household Waste. EU only.

Symbols

3

0088

Continued overleaf