- 6 -
Precautions before use
■
Power Requirement
●
The unit requires a power outlet with independently wired, 220-240V single-phase, 20A and grounding.
Note) Do not use a 2 pin plug adaptor and break a grounding pin of power cord. They may cause electric shock.
Location & Environment
●
Set up the unit on the center of a flat stable table.
Keep the unit away from walls and other equipment.
Note) Leave minimum of 20cm of space around the unit as well as 1m of space above the unit.
Do not place anything flammable near the unit. If not, conduct fire prevention.
Piping of Cooling Water, Vacuum Pump & Argon Gas
●
Connect cooling water and pressurized air to the unit as shown in figure below.
If necessary, connect argon gas to the unit.
Note) When the external chiller HPC-20 is not used for water supply, please keep cooling water temperature under 60
.
℃
Improper temperature may cause malfunction of the unit.
.
Note) The unit will not work without the proper connection of WATER A and WATER B
Note) Cut the blue piping tube for air & argon gas (5m) to proper length.
Note) Connect the white piping tube completely until the red line. Improper piping may cause water leaking.
Note) Connect CHILLER SIGNAL terminal completely with screws.
Note) Keep the hose length short to decrease the flow resistance within hose to a minimum.
Note) Keep the air pressure between 0.5 and 1.0MPa.
Note) Keep the argon gas pressure between 0.5 and 1.0MPa.
Note) Place the hose in a secure location where it will not get stepped on or tripped over.
The installation is finished.
CHILLER OUTLET
CHILLER
SIGNAL
ARGON
0.7-0.8MPa
WATER A WATER B
DRAIN COCK
Cut blue piping tube for use.
External compressed air 0.5~1.0Mpa
Power Cord
Cut blue piping tube for use.
Argon Gas 0.7~0.8Mpa
Connection for
optional chiller
(HPC-20)
Cooling water input
Connection for optional chiller (HPC-20)
or water connection part (DW-60)
Cooling water exhaust
Connection for optional chiller (HPC-20)
or water connection part (DW-60)
AIR IN
0.5-1.0MPa
EXHAUST
FILTER