KCI ActiV.A.C. Скачать руководство пользователя страница 1

For Clinician Use Only 

Clinician Reference Guide

ActiV.A.C.

®

Therapy System

Questions and Information

Clinician Feature Identification

Changing to Patient or Clinician Mode

Clinician Mode Home Screen

Settings

This is not intended as a comprehensive manual. For additional
information concerning proper use of the ActiV.A.C.

®

Therapy

System, consult the ActiV.A.C.

®  

User Manual (located in the

hard shell delivery case) and contact/or KCI.

If you are a patient, consult your treating physician and the
ActiV.A.C.

®

Therapy System User Manual concerning the proper

use of the ActiV.A.C.

®

Therapy System.

Touch Screen User Interface

KCI Authorized

Maintenance

Access Only

USB Data Port

(Clinician Use Only)

Power Connection

Battery Charging

Indicator Light

Power On/Off Button

Infrared Data Port

(Clinician Use Only)

Power the ActiV.A.C.

®

Therapy Unit On or Off

Press and hold         for approximately 2 seconds to turn the
ActiV.A.C.

®

Therapy Unit On or Off.

The ActiV.A.C.

®

Therapy Unit should be delivered to a new

patient already set to the Clinician Mode.

The ActiV.A.C.

®

Therapy Unit will return to the Patient Mode after

about 15 minutes of Touch Screen User Interface inactivity. 

ActiV.A.C.

®

Canister

All trademarks and service marks designated herein are property
of KCI Licensing, Inc., its affiliates, and licensors. This product is
patented and/or subject to pending patent. ©2010 KCI
Licensing, Inc. All rights reserved.
DSL#09-06-346 • REV7/09 • LIT#29-B-216

Help

Button

Indications, Contraindications, Warnings, Precautions

and other important Safety Information are contained in

the V.A.C.

®

Therapy General Safety Information Sheet

which should always be kept in the pocket inside the

front flap of the carrying case. If missing, contact

KCI Ireland at 1800 33 33 77.

Important Safety Information 

accompanies this device.

Therapy On/Off Button

Therapy Status Bar

and Display Area

Screen Guard

Mode Indicator 

Current Date/Current Time

This icon rotates when the
ActiV.A.C.

®

Therapy Unit is

applying negative pressure.

Battery Level Indicator This “plug” indicator

appears while plugged

into a wall outlet.

Therapy Start Screen

Press           to access the 

‘Help Menu’

.

Press 

‘Patient Mode’

or

‘Clinician Mode’

to go to 

appropriate mode.

One of two confirmation screens will appear, depending on which
mode is being accessed.

From Clinician Mode Home screen, press 

‘Therapy’

then 

‘Next’

, then 

‘Settings’

to access this screen.

Once the settings are chosen, this 

‘Confirm’

screen will appear.

Press 

‘OK’

to continue and                    to start therapy.

Press 

‘Settings’

to access and:

 Change 

‘Pressure’

settings              

 Change 

‘Intensity’

 Toggle between 

‘Continuous’

         

 Set 

‘Intermittent’

   and 

‘Intermittent’

therapy                  therapy times

 

‘Settings’

- Manually set therapy     

 

‘Seal Check

- Assists in finding leaks

 

‘Settings Guide’

- Guide for preset therapy settings

 

‘History’

- View or export therapy history

Press 

‘Exit’

to return to the Clinician Home screen.

Settings changed manually take immediate
effect when therapy is on.

From this screen, the 

‘Seal Check

tool can be used to view

integrity of V.A.C.

®

Dressings.

Press 

‘Patient Mode’

to

go to Patient Mode.

Press 

‘Clinician Mode’

to go to Clinician Mode.

Press and hold 

‘OK’

at least 5 seconds

to proceed to 

‘Clinician Mode’

.

Manual Therapy Settings

Manual Therapy Settings cont.

Log Tool

Seal Check

Tool