The medical device must be dry when transporting it to the site of re‐
processing.
▶
Do not place it in a solution or the like.
▶
Reprocess the medical device as soon as possible after treatment.
▶
Remove the tip from the SONICflex using the torque wrench.
6.2 Cleaning
CAUTION
Malfunctions from cleaning in an ultrasonic unit.
Defects in the product.
▶ Only clean manually or in a thermodisinfector.
Preparation methods according to ISO 17664
39