Instructions for use SMARTair Evo, SMARTair Evo mobile, SMARTair Evo FLEXspace
8 Safety checks in accordance with IEC 62638 (VDE 0701 / 0702)
8 Safety checks in accordance with IEC 62638
(VDE 0701 / 0702)
The applicable national legal regulations must be observed during the use of the
device, in particular:
▪ the applicable occupational health and safety regulations.
▪ the applicable accident prevention regulations.
The overarching guidelines and/or national laws, national regulations and the
rules of technology applicable to this product for start-up and use of the KaVo
product for the intended purpose are to be applied and complied with.
Compliance with the requirements of VDE 0701-0702 concerning product safety
upon start-up, after changes and repair as well as in recurrent tests is required.
Caution: Electrical safety checks have to be carried out by electrical engineering
professionals or by individuals trained in electrical engineering and supervised
by an electrical engineering professional.
An electrical engineering professional is defined to be anyone whose training,
knowledge and experience as well as knowledge of the pertinent regulations al-
low him or her to assess the imminent work and recognise possible hazards. It
is at the discretion of the electrical engineering professional in charge to decide
whether nor not further individual tests are required in order to attain the
safety target.
If safety deficiencies or damage is/are detected during the recurrent test, the
testing process must be discontinued and the equipment must be labelled ac-
cordingly and banned from further use or submitted for repair of the defect.
Annual tests
Note
According to § 53 Para. 2 of the Workplace Ordinance, vacuum devices must
be checked every 2 years for function by an expert. After consulting with the
Trade Supervisory Office in Celle, the test can be performed by a dental tech-
nician who has been trained how to use equipment in his journeyman train-
ing.
Note
KaVo recommends checking function annually and observing the sections
"Cleaning and servicing" and "Troubleshooting" in these instructions for use.
See also:
2
7 Cleaning and servicing, Page 37
See also:
2
▶
Check the filter for damage.
▶
Check the housing and hose lines for leaks and cracks.
▶
Check the function of the control device and the amount of filtration.
▶
Make sure that dust extraction system has original filters and all the neces-
sary filter.
▶
Mount the cover loosely, and check if it is sucked tight from the vacuum
when the unit is turned on. A flow of air must be detectable at the intake.
▶
The person performing the check must provide the date and signature.
41 / 46