10 Symbols and labeling
60
KaVo Scan eXam One
10.3 Warnings and precautions
THE UNIT IS A CLASS 1 LASER PRODUCT
NOTICE!
When covers are removed the unit is a
class 3B laser product – avoid exposure to the laser
beam.
Use of controls or adjustments or performance of
procedures other than those specified herein may
result in hazardous laser radiation exposure or
other non-compliance.
•
When handling imaging plates, protective
covers and hygiene bags always take the ap-
propriate hygiene measures and precautions
to prevent cross contamination. New protec-
tive cover must be used on every use.
•
The imaging plates are harmful if swallowed.
•
Do not move or knock the unit when it is pro-
cessing an imaging plate.
•
This unit must only be used to process image
plates supplied by the manufacturer and
must not be used for any other purpose.
•
NEVER use imaging plates, protective covers
or hygiene bags from other manufacturers.
•
This unit, or its accessories, must not be
modified, altered or remanufactured in any
way.
•
Only the manufacturer’s authorized service
personnel are authorized to carry out main-
tenance and repair of the unit. There are no
user serviceable parts inside the unit.
•
Unit is not suitable for use in the presence of
flammable anaesthetic mixture with air or
with oxygen or nitrous oxide.
•
In order to maintain safe and correct func-
tioning of the unit, only the power supply
unit (PSU) delivered with the unit or distrib-
uted by authorized dealers. Please refer to
the unit technical specifications for a list of
the PSUs.
•
For ethernet connections, use an unshielded
CAT6 RJ-45 LAN cable, so that multiple
chassis must not be connected! The PC /
Ethernet switch to which unit is connected to
must be class I or class II according IEC
60950. After installation check that the IEC
60601-1 leakage current levels are not ex-
ceeded.
•
If the PC/ Ethernet switch to which the de-
Approved:
Reviewed: Jalkanen Joni 2017-02-13 16:00
Reviewed
See PDM system to determine the status of this document. Printed out: 2017-03-17 10:35:17
D510414, 5
Copyright © by PaloDEx Group Oy. All rights reserved.