8 Other information
KaVo Scan eXam One
47
8 Other information
8.1 Quality control
To ensure maximum system performance
1. Observe “Exposure level” indication on the ap-
plication SW to see that the x-ray settings are
optimal.
2. Perform self quality control regularly according
instructions provided with quality control test
set SP00267
(Intra digi QC IEC phantom w. instructions).
8.2 Unit care
WARNING:
Switch the unit off and disconnect it from the main
power supply before cleaning or disinfecting the
unit. Do not allow liquids to enter the unit.
8.3 Unit cleaning
To clean the unit use a non abrasive cloth
moistened with:
•
cool or lukewarm water
•
soapy water
•
mild detergent
•
isopropyl alcohol
•
or ethanol (ethyl alcohol) 70 - 96%
•
CaviCide, CaviWipes by Metrex
•
FD322 by Dürr Dental
•
Easydes by Kiilto
After cleaning wipe the unit with a non abrasive
cloth moistened with water. Never use solvents or
abrasive cleaners to clean the unit. Never use
unfamiliar or untested cleaning agents. If you are
not sure what the cleaning agent contains, DO NOT
use it.
If you use a spray cleaning agent DO NOT spray it
directly into the unit door.