10. Place one to three pouches per tray or cassette.
11. Select an autoclave cycle with a minimum sterilisation temperature of
132° C (269° F).
NOTE
The minimum time and temperature values listed below reflect the
specific discrete values used to validate the sterilisation process for the
device.
•
Use distilled water to avoid reducing the tip life. The high
temperatures of the autoclave could cause the chlorine in tap water
to corrode the tip contacts.
•
For Pre-Vac autoclaves minimum sterilisation time is 3 minutes.
•
For Gravity autoclaves minimum sterilisation time is 10 minutes.
•
For any autoclave, use a minimum dry time of 30 minutes.
12. Whe the cycle has completed, store the tip for later use.
13.1.2 Storage
About this task
Follow these instructions to store cleaned scanner tips.
Steps
1. Wipe water off the mirror using non-woven optical wipes.
2. Optional: Insert tip into a sterilization pouch.
3. Store for later use.
What to do next
Prior to scanning, visually confirm the scanning mirror and scanner
connection points of the scanning tip are clean and dry. To clean either
surface, use an optical cloth or 2x2 gauze with a small amount of alcohol to
gently remove residue. Ensure each surface is dry before connecting it to the
scanner.
13 Cleaning / preventive maintenance
Instructions for use
KaVo ProXam iOS 103