7 Maintenance
7
Maintenance
The maintenance and calibration procedure intervals described here are minimum requirements and
recommendations. The maintenance and calibration procedures can be made more frequent and
stringent to comply with local regulations regarding the use and maintenance of dental X-ray devices.
7.1
Cleaning and decontamination
NOTICE!
Decontamination techniques for the device, its accessories and the room must comply
with all laws and regulations within the local jurisdiction.
NOTICE!
Clean the patient positioning accessories before their first use.
CLEANING THE SURFACES THAT THE PATIENT TOUCHES:
All surfaces and parts that the patient touches or comes into contact with, must be decontaminated
after each patient. Use a decontaminant that is formulated specifically for decontaminating dental
equipment and use it in accordance with the instructions supplied with it. Wipe all items and surfaces
dry after the decontamination.
WARNING!
Do not use any disinfecting aerosols, since the vapors could ignite causing injuries.
NOTICE!
Wear gloves and other protective equipment during the decontamination process.
Autoclave
Lip Supports and Bite Blocks may be autoclaved in 121°C (250 F) temperature for max. 3 minutes. Other
parts are not autoclavable.
NOTICE!
Autoclaving is not recommended as cleaning method, and the parts are not intended to
be sterilized.
CLEANING THE DEVICE:
The device should be cleaned regularly.
CAUTION!
Switch the device off or isolate it from mains before cleaning the device.
CAUTION!
Do not allow water or other cleaning liquids to enter the device's interior since it might
cause short-circuits or corrosion. If you use a spray cleaner do not spray into any ventilation
grills.
CAUTION!
Clean the dust off the device regularly. The device might overheat if excess dust is
gathered on the ventilation grills.
Device surfaces
All device surfaces can be wiped clean with a soft cloth dampened with a mild detergent, for example
soapy water. DO NOT use abrasive cleaning agents or polishes.
Patient positioning panel
Wait for the panel to dry or wipe it dry before use. Moisture can affect the usability of the panel.
Allowed cleaning agents for cleaning and decontamination of the device:
• Distilled water
• Ethanol 96%
• Isopropyl alcohol
• Soapy water
• CaviCide
™
, CaviWipes
™
or Metasys
™
disinfectant
70
ORTHOPANTOMOGRAPH
™
OP 3D