4 Firmware description
82
4.2 Detailed View
R3220 CPU
Executes the main and boot SW of the unit.
Boot software
Loads the main software into the RAM and starts it.
Handles the initial unit software loading and boot mode
updates.
Main software
Controls and coordinates unit movements, exposure,
image acquisition and transfer.
The image data acquired during image capture are stored
in the internal memory of the unit. After image acquisition
the image data are transferred to the workstation using the
external 1000 Base - T Ethernet UDP/IP Ethernet.
It also handles multi connect and up to five clients (drivers,
S2Terminals or replicate PC touch screen displays) can be
connected to the unit via the external Ethernet.
It also offers firmware upgrade services for itself and
software upgrade for touch screen display. It also offers
core (FW) update service for R3300 IO1, R3400 IO2 and
R3210 Ceph as there is no direct connection from the
S2Terminal to these three boards.
R3300 IO1 software
Controls non-critical movements
R3400 IO2 software
Controls non-critical movements, IO, lights etc.
Communicates with CPU over a 100Base-T interface.
Touch screen display software
Holds the unit calibration data that is transfered to the
R3220 CPU during start up.
R3210 Ceph CPU
Controls the cephalometric imaging movements on the
ceph end (secondary collimator and sensor movements).
Controls cephalometric image acquisition.
Содержание ORTHOPANTOMOGRAPH OP 3D Pro
Страница 1: ...ORTHOPANTOMOGRAPH OP 3D Pro 3D Dental X ray System Service Manual ENGLISH 216508 rev 1 0 805 5127...
Страница 2: ......
Страница 4: ......
Страница 18: ...1 Introduction 10...
Страница 22: ...2 Unit description 14 Panoramic lights 1 Midsagittal light 2 FH light 1 Image layer 2 TMJ light...
Страница 28: ...2 Unit description 20 2 6 Unit movements Panoramic unit movements R C J D...
Страница 31: ...2 Unit description 23...
Страница 32: ...2 Unit description 24...
Страница 37: ...3 Electrical description 29 Circuit boards Cephalometric unit...
Страница 54: ...46...
Страница 87: ...79...
Страница 88: ...80...
Страница 96: ...4 Firmware description 88...
Страница 101: ...5 Hardware operation 216508 rev 1 NEW KAVO 93 5 6 Z movement 2...
Страница 105: ...5 Hardware operation 216508 rev 1 NEW KAVO 97 5 10 Secondary collimator movement 2...
Страница 106: ...5 Hardware operation 98 NEW KAVO 216508 rev 1 5 11 Ceph sensor movement 2...
Страница 107: ...5 Hardware operation 216508 rev 1 NEW KAVO 99 5 12 Imaging chain 5 12 1 3D imaging chain...
Страница 108: ...5 Hardware operation 100 NEW KAVO 216508 rev 1 5 12 2 Panoramic imaging chain...
Страница 109: ...5 Hardware operation 216508 rev 1 NEW KAVO 101 5 12 3 Cephalometric imaging chain...
Страница 110: ...5 Hardware operation 102 NEW KAVO 216508 rev 1 5 12 4 Exposure control...
Страница 112: ...5 Hardware operation 104 NEW KAVO 216508 rev 1...
Страница 122: ...6 The s2terminal and service functions 114...
Страница 192: ...8 Maintenance procedure 184...