7 Error codes and troubleshooting
172
Actuator malfunction can be caused either by a faulty
motor or faulty driving electronics. Use the s2terminal, the
drivemotor
command, with suitable parameters, to check
the status of the motors. If there is no motion the faulty
component can be easily found by replacing them one by
one, the motor, the driving electronics and then the cabling
until the faulty component is identified.
Sensor malfunctions can be solved in same way as
actuator malfunctions. Selecting service mode and then
using the
systemstatus
command will show the status of
most sensors (optosensor and micro switches). Trouble
shoot by manually moving the unit and following the status
changes for the sensor you wish to check. Alternatively
start the
drivemotor
command in manual mode for
example
drivemotor X man
will drive the X motor back
and forth (when chin support up/down keys are pressed). If
the status does not change when an opto or mechanical
switch is activated, then changing the sensor or its related
cabling should fix the problem.
If related cables and switches are found to be working, the
problem is probably caused by a faulty logic board (R3220,
R3300 and/or R3400). The related block diagrams give an
overview of what components comprise a particular
motion.
Note!
Some of sensor signals travel through the serial in-
terface (J3213
JG205). If such a signal is shown incor-
rectly, the fault may be in either the cabling, the sensor, or
the serial interface board (EG200).
7.6.10
Z-Motion control
7.6.10.1
Main support up/down (Z) movement does
not work
•
Check that the LEDs UP/DOWN and ENA on
R4300 come on when the Z up/down buttons are
pressed. If not, check that both direction and ena-
ble signals from the buttons are reaching the
R3400 board by using service command
system-
status
. Check the connection of the buttons and
replace if needed.
•
Check that the voltage selector on EA200 is set to
the mains voltage being used.
•
Check fuse 1 on R3400.
Replace if blown.
Содержание ORTHOPANTOMOGRAPH OP 3D Pro
Страница 1: ...ORTHOPANTOMOGRAPH OP 3D Pro 3D Dental X ray System Service Manual ENGLISH 216508 rev 1 0 805 5127...
Страница 2: ......
Страница 4: ......
Страница 18: ...1 Introduction 10...
Страница 22: ...2 Unit description 14 Panoramic lights 1 Midsagittal light 2 FH light 1 Image layer 2 TMJ light...
Страница 28: ...2 Unit description 20 2 6 Unit movements Panoramic unit movements R C J D...
Страница 31: ...2 Unit description 23...
Страница 32: ...2 Unit description 24...
Страница 37: ...3 Electrical description 29 Circuit boards Cephalometric unit...
Страница 54: ...46...
Страница 87: ...79...
Страница 88: ...80...
Страница 96: ...4 Firmware description 88...
Страница 101: ...5 Hardware operation 216508 rev 1 NEW KAVO 93 5 6 Z movement 2...
Страница 105: ...5 Hardware operation 216508 rev 1 NEW KAVO 97 5 10 Secondary collimator movement 2...
Страница 106: ...5 Hardware operation 98 NEW KAVO 216508 rev 1 5 11 Ceph sensor movement 2...
Страница 107: ...5 Hardware operation 216508 rev 1 NEW KAVO 99 5 12 Imaging chain 5 12 1 3D imaging chain...
Страница 108: ...5 Hardware operation 100 NEW KAVO 216508 rev 1 5 12 2 Panoramic imaging chain...
Страница 109: ...5 Hardware operation 216508 rev 1 NEW KAVO 101 5 12 3 Cephalometric imaging chain...
Страница 110: ...5 Hardware operation 102 NEW KAVO 216508 rev 1 5 12 4 Exposure control...
Страница 112: ...5 Hardware operation 104 NEW KAVO 216508 rev 1...
Страница 122: ...6 The s2terminal and service functions 114...
Страница 192: ...8 Maintenance procedure 184...