7 Error codes and troubleshooting
171
7.6.9
Motion control
There are five possible types of malfunction:
•
Mechanical malfunction
•
Power supply problem
•
Actuator malfunction
•
Sensor malfunction or
•
Control malfunction
Mechanical malfunctions are easy to identify because in
most of cases they are clearly visible. For this reason they
are out of the scope of this document.
In most cases it is good practice to check the condition of
power supply first. The trouble shooting related to the
power distribution is described in chapters 7.6.2 and 7.6.3.
Basically checking whether the power supply indicator
LEDs are on or off will indicate the condition of the power
distribution.
If voltage is not 0.92-0.93V check the
condition of the cable.
If the cable is faulty, replace it.
If the cable is OK, replace
EG200.
If voltage is near 0.92-0.93V, measure the volt-
age from TP7 (kV feedback from tube).
If voltage is 1.25-1.26V measure voltage
from TP13 (mA feedback).
If voltage is not 0.58V replace
R3810.
If replacing R3810 does
not solve the problem, re-
place the tubehead.
If voltage is 0.58V check condi-
tion of the ribbon cable.
If faulty, replace it.
If the cable is OK, replace
EG200.
Содержание ORTHOPANTOMOGRAPH OP 3D Pro
Страница 1: ...ORTHOPANTOMOGRAPH OP 3D Pro 3D Dental X ray System Service Manual ENGLISH 216508 rev 1 0 805 5127...
Страница 2: ......
Страница 4: ......
Страница 18: ...1 Introduction 10...
Страница 22: ...2 Unit description 14 Panoramic lights 1 Midsagittal light 2 FH light 1 Image layer 2 TMJ light...
Страница 28: ...2 Unit description 20 2 6 Unit movements Panoramic unit movements R C J D...
Страница 31: ...2 Unit description 23...
Страница 32: ...2 Unit description 24...
Страница 37: ...3 Electrical description 29 Circuit boards Cephalometric unit...
Страница 54: ...46...
Страница 87: ...79...
Страница 88: ...80...
Страница 96: ...4 Firmware description 88...
Страница 101: ...5 Hardware operation 216508 rev 1 NEW KAVO 93 5 6 Z movement 2...
Страница 105: ...5 Hardware operation 216508 rev 1 NEW KAVO 97 5 10 Secondary collimator movement 2...
Страница 106: ...5 Hardware operation 98 NEW KAVO 216508 rev 1 5 11 Ceph sensor movement 2...
Страница 107: ...5 Hardware operation 216508 rev 1 NEW KAVO 99 5 12 Imaging chain 5 12 1 3D imaging chain...
Страница 108: ...5 Hardware operation 100 NEW KAVO 216508 rev 1 5 12 2 Panoramic imaging chain...
Страница 109: ...5 Hardware operation 216508 rev 1 NEW KAVO 101 5 12 3 Cephalometric imaging chain...
Страница 110: ...5 Hardware operation 102 NEW KAVO 216508 rev 1 5 12 4 Exposure control...
Страница 112: ...5 Hardware operation 104 NEW KAVO 216508 rev 1...
Страница 122: ...6 The s2terminal and service functions 114...
Страница 192: ...8 Maintenance procedure 184...