7 Error codes and troubleshooting
169
7.6.8.2
Images too dark or too light (Wrong dose
generated)
•
Check the condition of the power supply by follow-
ing trouble shooting guide 7.6.2.3 for no X-rays.
•
Check that X-ray unit calibrations have been car-
ried out.
Preheat check
•
Set the X-ray unit to the 3D imaging mode (and us-
ing the s2terminal type patient), adjust the mA set-
tings from the GUI and then measure the
respective voltages from TP11 (Preheat refer-
ence).
mA check
•
Set the X-ray unit to the 3D imaging mode (and us-
ing the s2terminal type
patient
), adjust the mA set-
tings from the GUI and measure the respective
voltages from TP11 (mA reference).
If the FAIL LED comes on
replace R3810. If the LED
still comes on after re-
placing R3810, replace
the tubehead.
If voltages differ from those listed below, replace
EG200.
8 mA
1.54V
10 mA
1.59V
12.5 mA
1.65V
15 mA
1.70V
If the voltages do not differ, set the mA to 15 mA
and measure voltage from TP15 (preheat feed-
back) when an exposure is taken.
if voltage is not 0.58V then replace
R3810.
If this does not solve the problem
still, replace the tubehead.
Содержание ORTHOPANTOMOGRAPH OP 3D Pro
Страница 1: ...ORTHOPANTOMOGRAPH OP 3D Pro 3D Dental X ray System Service Manual ENGLISH 216508 rev 1 0 805 5127...
Страница 2: ......
Страница 4: ......
Страница 18: ...1 Introduction 10...
Страница 22: ...2 Unit description 14 Panoramic lights 1 Midsagittal light 2 FH light 1 Image layer 2 TMJ light...
Страница 28: ...2 Unit description 20 2 6 Unit movements Panoramic unit movements R C J D...
Страница 31: ...2 Unit description 23...
Страница 32: ...2 Unit description 24...
Страница 37: ...3 Electrical description 29 Circuit boards Cephalometric unit...
Страница 54: ...46...
Страница 87: ...79...
Страница 88: ...80...
Страница 96: ...4 Firmware description 88...
Страница 101: ...5 Hardware operation 216508 rev 1 NEW KAVO 93 5 6 Z movement 2...
Страница 105: ...5 Hardware operation 216508 rev 1 NEW KAVO 97 5 10 Secondary collimator movement 2...
Страница 106: ...5 Hardware operation 98 NEW KAVO 216508 rev 1 5 11 Ceph sensor movement 2...
Страница 107: ...5 Hardware operation 216508 rev 1 NEW KAVO 99 5 12 Imaging chain 5 12 1 3D imaging chain...
Страница 108: ...5 Hardware operation 100 NEW KAVO 216508 rev 1 5 12 2 Panoramic imaging chain...
Страница 109: ...5 Hardware operation 216508 rev 1 NEW KAVO 101 5 12 3 Cephalometric imaging chain...
Страница 110: ...5 Hardware operation 102 NEW KAVO 216508 rev 1 5 12 4 Exposure control...
Страница 112: ...5 Hardware operation 104 NEW KAVO 216508 rev 1...
Страница 122: ...6 The s2terminal and service functions 114...
Страница 192: ...8 Maintenance procedure 184...