7 Error codes and troubleshooting
168
7.6.8
X-ray generation
7.6.8.1
No X-rays generated
•
Check condition of the power supply by following
trouble shooting guide 7.6.2.3 for no X-rays.
•
Check the condition of the exposure switch by fol-
lowing trouble shooting guide 7.6.7.2 for Exposure
sequence starts but no X-rays.
•
Set the X-ray unit to the imaging mode (CT or Pan),
disable movements (s2terminal), press exposure
switch and check the status of the Preheat LED
(D3) on the R3810.
If the voltage goes above 2V re-
place R3300.
If voltage is near 0V then meas-
ure voltage at connector J3202
pin 7.
If voltage goes above 2V
check and if faulty replace
cable between J3303 and
J3202.
If voltage is near 0V re-
place R3220.
If LED D3 does not come on, check the condition
of the cable between connectors J3314 and
J3810.
If cable faulty replaced it.
If cable OK replace R3810.
If LED D3 comes on, check the status of the ex-
posure LED (D31) on the R3810.
If LED D31 does not come on, check the
condition of the cable between connec-
tors JG214 and J3810.
If cable faulty, replace it.
If cable OK, check if the FAIL
LED is on.
If not replace R3810.
Содержание ORTHOPANTOMOGRAPH OP 3D Pro
Страница 1: ...ORTHOPANTOMOGRAPH OP 3D Pro 3D Dental X ray System Service Manual ENGLISH 216508 rev 1 0 805 5127...
Страница 2: ......
Страница 4: ......
Страница 18: ...1 Introduction 10...
Страница 22: ...2 Unit description 14 Panoramic lights 1 Midsagittal light 2 FH light 1 Image layer 2 TMJ light...
Страница 28: ...2 Unit description 20 2 6 Unit movements Panoramic unit movements R C J D...
Страница 31: ...2 Unit description 23...
Страница 32: ...2 Unit description 24...
Страница 37: ...3 Electrical description 29 Circuit boards Cephalometric unit...
Страница 54: ...46...
Страница 87: ...79...
Страница 88: ...80...
Страница 96: ...4 Firmware description 88...
Страница 101: ...5 Hardware operation 216508 rev 1 NEW KAVO 93 5 6 Z movement 2...
Страница 105: ...5 Hardware operation 216508 rev 1 NEW KAVO 97 5 10 Secondary collimator movement 2...
Страница 106: ...5 Hardware operation 98 NEW KAVO 216508 rev 1 5 11 Ceph sensor movement 2...
Страница 107: ...5 Hardware operation 216508 rev 1 NEW KAVO 99 5 12 Imaging chain 5 12 1 3D imaging chain...
Страница 108: ...5 Hardware operation 100 NEW KAVO 216508 rev 1 5 12 2 Panoramic imaging chain...
Страница 109: ...5 Hardware operation 216508 rev 1 NEW KAVO 101 5 12 3 Cephalometric imaging chain...
Страница 110: ...5 Hardware operation 102 NEW KAVO 216508 rev 1 5 12 4 Exposure control...
Страница 112: ...5 Hardware operation 104 NEW KAVO 216508 rev 1...
Страница 122: ...6 The s2terminal and service functions 114...
Страница 192: ...8 Maintenance procedure 184...