6 The s2terminal and service functions
107
6.2 s2terminal commands
These commands allow images to be downloaded directly
from the unit memory and the unit reserved if more than
one client (an s2terminal or driver in the same PC or
another PC) is connected to the unit. Additional commands
(under help) allow firmware to be updated.
---- S2 Terminal Help --------------------------------------
------------------------------------------------------------------
Downloading raw images - xi
1. Disable (no image capture) or close the dental imaging
program in the PC(s) connected to the unit to ensure that
the s2terminal has reserved the unit.
2. In the main
s2terminal
folder create a subfolder named
imageseries
.
3. Key in
xi
and press
ENTER
.
The last image taken by the unit will be downloaded from
the unit memory into the folder.
Image will be saved in DCM (DICOM) format. The saved
image can be opened with the dental imaging software or
image viewers such as
ImageJ
or
IrfanView
.
Note!
If the unit was switched off after the last image was
taken the image will be lost.
Reserving the unit - xr
If you wish to take an image with the dental imaging
software while the s2terminal is running you must release
the s2terminal and reserve the dental imaging software.
1. Key in
xr
and then press
ENTER
.
------------------------------------------------------------------
xh
help
xi
receive image from the device
xr
reserve device
xq
quit
xr
Содержание ORTHOPANTOMOGRAPH OP 3D Pro
Страница 1: ...ORTHOPANTOMOGRAPH OP 3D Pro 3D Dental X ray System Service Manual ENGLISH 216508 rev 1 0 805 5127...
Страница 2: ......
Страница 4: ......
Страница 18: ...1 Introduction 10...
Страница 22: ...2 Unit description 14 Panoramic lights 1 Midsagittal light 2 FH light 1 Image layer 2 TMJ light...
Страница 28: ...2 Unit description 20 2 6 Unit movements Panoramic unit movements R C J D...
Страница 31: ...2 Unit description 23...
Страница 32: ...2 Unit description 24...
Страница 37: ...3 Electrical description 29 Circuit boards Cephalometric unit...
Страница 54: ...46...
Страница 87: ...79...
Страница 88: ...80...
Страница 96: ...4 Firmware description 88...
Страница 101: ...5 Hardware operation 216508 rev 1 NEW KAVO 93 5 6 Z movement 2...
Страница 105: ...5 Hardware operation 216508 rev 1 NEW KAVO 97 5 10 Secondary collimator movement 2...
Страница 106: ...5 Hardware operation 98 NEW KAVO 216508 rev 1 5 11 Ceph sensor movement 2...
Страница 107: ...5 Hardware operation 216508 rev 1 NEW KAVO 99 5 12 Imaging chain 5 12 1 3D imaging chain...
Страница 108: ...5 Hardware operation 100 NEW KAVO 216508 rev 1 5 12 2 Panoramic imaging chain...
Страница 109: ...5 Hardware operation 216508 rev 1 NEW KAVO 101 5 12 3 Cephalometric imaging chain...
Страница 110: ...5 Hardware operation 102 NEW KAVO 216508 rev 1 5 12 4 Exposure control...
Страница 112: ...5 Hardware operation 104 NEW KAVO 216508 rev 1...
Страница 122: ...6 The s2terminal and service functions 114...
Страница 192: ...8 Maintenance procedure 184...