KaVo MASTERtorque Mini LUX M8700 L Скачать руководство пользователя страница 8

Instructions for use MASTERtorque Mini LUX M8700 L - 3.001.0000MASTERtorque Mini LUX M8700 LK -
3.001.3400MASTERtorque Mini LUX M8700 LS - 3.001.0600

2 Safety | 2.3 Accessories and combination with other equipment

8 / 30

2.3 Accessories and combination with other equipment

Use of un-authorised accessories or un-authorised modifications of the device
could lead to injury.

Only use accessories that have been approved for combination with the
product by the manufacturer.

Only use accessories that are equipped with standardised interfaces.

Do not make any modifications to the device unless these have been ap-
proved by the manufacturer of the product.

2.4 Qualification of personnel

Application of the product by users without the appropriate medical training
could injure the patients, the users or third parties.

Make sure that the user has read and understood the instructions for use.

Only employ the device if the user has the appropriate medical training.

Observe national and regional regulations.

The improper use of the device could lead to burns or injuries.

Never touch soft tissue with the handpiece head or instrument cover.

Do not use the medical device as a light probe.

Use an appropriate light probe for illumination of the oral cavity or site of
preparation.

After treatment, place the medical device properly in the holder, without the
cutter or grinder.

2.5 Service and repair

Repairs, servicing and safety checks may only be performed by trained service
personnel. The following persons are authorised to do this:

▪ Service technicians of KaVo branches after the appropriate product training

▪ Service technicians of KaVo authorised dealers after the appropriate product

training

Observe all the following items during servicing work:

Have the service and testing tasks carried out according to the Medical
Product Operator Ordinance.

After servicing, interventions, and repairs of the device and before re-use,
the device must be subjected to safety checks by the service personnel.

Following expiry of the warranty, check the cutter/grinder holding system
and quick stop once a year.

KaVo recommends specifying in-house service intervals where the medical
device is brought to a professional shop for cleaning, servicing and a func-
tion check. Defined the service interval depending on the frequency of use.

Содержание MASTERtorque Mini LUX M8700 L

Страница 1: ...Instructions for use MASTERtorque Mini LUX M8700 L 3 001 0000 MASTERtorque Mini LUX M8700 LK 3 001 3400 MASTERtorque Mini LUX M8700 LS 3 001 0600...

Страница 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Phone 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer KaVo Dental GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Страница 3: ...ng tool or diamond grinder 15 6 Troubleshooting 16 6 1 Changing the O rings on the MULTIflex LUX MULTIflex LED coupling M8700 L 16 6 2 Cleaning the spray nozzle 16 6 3 Changing the water filter 16 7 R...

Страница 4: ...ge 25 8 Optional aids 26 9 Terms and conditions of warranty 27 Instructions for use MASTERtorque Mini LUX M8700 L 3 001 0000MASTERtorque Mini LUX M8700 LK 3 001 3400MASTERtorque Mini LUX M8700 LS 3 00...

Страница 5: ...ce For repairs please contact your local dealer or the KaVo Repair Service directly 49 0 7351 56 1900 service reparatur kavokerr com Target group This document is intended for dentists and their assis...

Страница 6: ...nditions Air pressure Transportation and storage conditions Humidity Protect from moisture Protect from impact HIBC Code Hazard levels The warning and safety notes in this document must be observed to...

Страница 7: ...ust be appropriately repro cessed or sterilised In the case of injury to soft tissue do not continue treatment in the oral cavity with compressed air driven instruments 2 2 Technical condition A damag...

Страница 8: ...juries Never touch soft tissue with the handpiece head or instrument cover Do not use the medical device as a light probe Use an appropriate light probe for illumination of the oral cavity or site of...

Страница 9: ...he following applications Removal of carious material cavity and crown preparations removal of fillings processing of tooth and restora tion surfaces A medical device according to relevant national st...

Страница 10: ...X MULTIflex LED couplings 3 3 Technical Specifications M8700 LS Drive pressure 2 6 to 3 0 bar 38 to 44 psi Recommended drive pressure 2 8 bar 41 psi Return air pressure 0 5 bar 7 psi Spray water press...

Страница 11: ...rding to EN ISO 7494 2 4 1 Mounting the MULTIflex LUX MULTIflex LED coupling to M8700 L WARNING Detachment of the medical device during treatment A medical device that is not properly locked in place...

Страница 12: ...coupling and the medical device and check the following pressures Drive pressure recommended drive pressure return air pressure spray water pressure and spray air pressure See also 2 3 2 Technical Sp...

Страница 13: ...ts service life Make sure that the turbine is accurately coupled and firmly seated on the coupling Mount the medical device accurately on the MULTIflex LUX MULTIflex LED coupling and push it backward...

Страница 14: ...tment Never use dental burs or diamond grinders with damaged or worn shafts CAUTION Danger of injury from cutters or grinders Infections or cuts Wear gloves or finger stalls CAUTION Hazard from defect...

Страница 15: ...utton while the dental bur or diamond grinder is ro tating Do not touch the dental bur or diamond grinder while it is rotating Never touch soft tissue with the handpiece head or instrument cover Remov...

Страница 16: ...between your fingers to form a loop Push the O ring to the front and remove it Insert new O rings into the grooves 6 2 Cleaning the spray nozzle WARNING Hazard from nonsterile products Infection dang...

Страница 17: ...overheat and damage the tooth Check the filter and exchange if necessary Check the spray water channels and clean the spray nozzles with the nozzle needle Mat no 0 410 0921 if necessary Unscrew the s...

Страница 18: ...essing To minimise the risk of infection during reprocessing always wear protect ive gloves Remove the tool from the medical device Remove all residual cement composite or blood immediately Do not pla...

Страница 19: ...ng can only be carried out with washer disin fectors in accordance with EN ISO 15883 1 7 2 3 Manual external disinfection WARNING Incomplete disinfection Infection hazard Only use disinfection procedu...

Страница 20: ...atibility of materials e g WL cid made by ALPRO Cover the medical device with the KaVo CLEANpac bag and place it on the corresponding care adapter Hold the can vertically Press the spray key for at le...

Страница 21: ...cess this medical device in an ultrasonic device Malfunction and material damage Clean manually or in a washer disinfector only 7 3 1 Automated internal and external cleaning and internal and external...

Страница 22: ...Spray KaVo recommends servicing the product as part of the reprocessing after each use i e after each cleaning disinfection and before each sterilisation Remove the tool from the medical device Cover...

Страница 23: ...e PLUS and attach it to coupling service point four on the far right A MULTIflex adaptor must be mounted there Press the handpiece together with the guide bush of the chuck to be ser viced against the...

Страница 24: ...as part of the reprocessing after each use i e after each cleaning disinfection and before each sterilisation Remove the tool from the medical device Service the product in the QUATTROcare See also 2...

Страница 25: ...p to 138 280 4 F Sterilisation parameters Select a suitable procedure depending on the available autoclave from the fol lowing sterilisation processes Steriliser with triple pre vacuum at least 3 minu...

Страница 26: ...Wrench for water filter 1 002 0321 Instrument stand 2151 0 411 9501 Nozzle pin 0 410 0921 Cleanpac 10 units 0 411 9691 MULTIflex spray head nozzle 0 411 9921 Spray head SIRONA for KaVo Spray 1 005 836...

Страница 27: ...ry KaVo shall not be liable for defects and their consequences that have arisen or may arise from natural wear improper handling cleaning or maintenance non compliance with operating maintenance or co...

Страница 28: ......

Страница 29: ......

Страница 30: ...1 011 1220 bd 20211130 03 en...

Отзывы: