9
Assembly instructions
9.1
Floor composition
The quality of the floor (screed) must comply with the load bearing capacity for
buildings
DIN 1055 page 3 , and the pressure resistance must correspond to DIN
18560 T1 .
The KEY Laser 3+ / III upgraded is delivered from the factory with soft castors.
Note
The castors may not be used on soft floors (rugs, etc.).
The screed, floor covering, adhesive and floor sealant must be appropriate for
castors.
KaVo assumes no liability for any damage that may occur if the castors are im‐
properly used on different floor coverings.
9.2
Media
9.2.1
Spray water
The KEY Laser 3+ / III upgraded has an internal spray water supply.
The spray water storage bottle may be filled only with distilled or demineralised water
(neutral pH).
See also: 4.2 Fill the spray water, Page 30
9.2.2
Spray air
A compressor is integrated in the KEY Laser 3+ / III upgraded to generate a system
air supply.
9.3
Electrical connection
CAUTION
Damage to the unit from wrong fuses or supply cables.
Damage to the unit.
▶ Provide separate fuses for the KEY Laser 3+ / III upgraded.
▶ Use a separate supply cable for the KEY Laser 3+ / III.
9.3.1
Customer
▪ The connection must correspond to the provisions of VDE 0100 for Schuko
sockets.
▪ Connection for USA 125V/30 AMP corresponding to NEMA 5-30R
▪ 230 V fuses with B16 or C16 m.c.b.
▪ 100-127 V fuses with 25 A
▪ 100-127 V fuses with 20 A, for Japan only
Instructions for use KEY Laser 3+ 1343 KEY Laser III 1243 upgraded
9 Assembly instructions | 9.1 Floor composition
73/76
Содержание KEY Laser 3+ 1343
Страница 1: ...Instructions for use KEY Laser 3 1343 KEY Laser III 1243 upgraded Always be on the safe side ...