▶ Document the measured value in the medical device reference.
6.6
Additional tests
▪ Visual check of the device and accessories
▪ Function check
▪ No visible leaks in the unit
▪ Check internal safety devices
– Emergency-off switch
– Key-operated switch
– Safety switch for swing arm (shutter locking)
– Safety switch for laser hose coupling
– Connection for external door contact
– Connection for external warning lamp
– Monitoring the cooling system
– Energy monitoring/control
– Shutter/absorber
– Safety foot control
▪ Testing of external safety devices
– Warning signs
Check if all required warning signs are affixed to all the access doors to the
KEY Laser 3+ / III upgraded treatment room.
– Warning lamps
Check if all the access doors to the KEY Laser 3+ / III upgraded treatment
room have warning lights.
– Door switch
If door switches have been installed, their function should be tested. When
a door is open, the KEY Laser 3+ / III upgraded may not switch to READY
mode.
– Safety goggles
All safety goggles must be checked for suitability.
Instructions for use KEY Laser 3+ 1343 KEY Laser III 1243 upgraded
6 Safety checks | 6.6 Additional tests
64/76