Safety checks
CAUTION
Damage due to improper servicing
Using improper servicing tools can restrict the function or damage the unit.
▶ Use at least the basic KaVo service equipment.
▶ After first start up, servicing tasks and manipulating safety-relevant compo‐
nents, run a safety check.
▶ Do a safety check at least every 12 months.
Combo recommends annual service checks before the safety checks.
▶ Observe the section on safety.
6.1
Scope of tests
The following tests must be carried out taking into account IEC 60601-1 and IEC
60825-1 and the regulations of the professional associations for safety and health
at the workplace (laser safety BGV B2).
The test runs according to the following table:
Frequency in Hz
Pulse energy in mJ
4, 6, 10 and 13
300
15, 20 and 25
200
S-PULSE
Frequency in Hz
Pulse energy in mJ
10, 13, 15, 20, 25 and 30
200
6.2
Safety check of energy
If the energy safety check is activated, in the following test steps are automatically
performed by the KEY Laser 3+ / III upgraded:
▪ The KEY Laser 3+ / III upgraded automatically runs a test of the average, max‐
imum, minimum, standard deviation and error (average in comparison to the
setpoint).
– The error may not /- 10%.
▪ The test results must be transmitted from the KEY Laser 3+ / III upgraded to the
terminal program. The report must be printed out and added to the medical de‐
vice reference and customer service report.
See also: 6.7 Safety check report (example), Page 65
Instructions for use KEY Laser 3+ 1343 KEY Laser III 1243 upgraded
6 Safety checks | 6.1 Scope of tests
62/76