Technician's instructions KaVoLUX 540 LED U / T
11 Safety checks - Test instructions | 11.2 Ceiling mounting
Scan the light head of the treatment light KaVoLUX 540 LED with the
probe
① Fastening screw of the handle support
when the gripping sleeve has been re-
moved
Measuring the equivalent unit leakage current
Threshold:
< 10 mA (maximum value!)
ACP
Protection class 1
WARNING
Electrical power.
Death or injury from electric shock.
▶ Conduct test for leakage current in devices of Protection Class 1 only after the
protective earth test has been passed.
WARNING
Electrical power.
Death or injury from electric shock.
▶ Prior to connecting the treatment centre to the sight window, disconnect the treat-
ment unit from the mains supply network.
Functional tests
The following conditions must be fulfilled in all function tests:
▪ The basic functions of the operating light must be guaranteed.
▪ The treatment light must be fit for use.
51 / 56