Care instructions Primus 1058 Life
2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit
18 / 44
2.4.6 Servicing the turbine return air filter
Cleaning the turbine return air filter
Note
The liquid collector must be checked weekly.
▶
If there is any liquid in the filter, drain the liquid and replace the O-rings of
the multiflex couplings.
▶
If necessary, replace the turbine return air filter (
Mat. no. 0.200.3098
)
and clean the filter holder.
Note
The turbine return air filter is in the same position in the dentist element vari-
ant with S table.
2.4.7 Reprocessing KaVo ERGOcam One
Note
The camera handle cannot be sterilised or disinfected with heat.
Cleaning
Note
Disinfect the camera before cleaning.
Disinfect after cleaning if necessary.
Exterior cleaning by hand
NOTICE
Damage to the camera due to improper cleaning.
Destroys the surface or the optical system of the camera.
▶
Do not use abrasives or aggressive cleansers!
▶
Do not immerse the camera in liquids!
▶
Switch main device switch off.
▶
Remove the camera.