Care instructions Primus 1058 Life
2 Setup methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist unit
2.4.3 Reprocessing of the instrument arm S table
Note
The instrument arm S table (
Mat. no. 1.007.4756
) can be removed to facil-
itate cleaning.
Cleaning and disinfection of the instrument arm S table
NOTICE
Damage caused by treatment in a washer disinfector.
▶
Do not sterilise the instrument arm S table in a washer disinfector.
Note
Neither the rinsing program nor the intensive germ reduction can be carried
out when the instrument arm is removed.
▶
Switch off the unit at the main switch.
▶
Attach the germ reduction attachment.
with patient part assembly kit / without patient part assembly kit
▶
Take the handpieces off the hoses.
13 / 44