The program settings and cleansers and disinfectants that must be
used can be found in the instructions for use of the thermodisinfector.
▶
Directly after automated cleaning, treat the medical device with the
care products and systems provided by KaVo.
6.2.3 Cleaning: Manual cleaning - internal
For a medical device with KaVo CLEANspray and KaVo DRYspray.
▶
Cover the medical device with the KaVo CLEANpac bag, and place it
on the corresponding care adapter. Press the spray button three times
for 2 seconds each time. Remove the medical device from the spray
attachment and let the cleaner work for one minute.
▶
Then rinse for 3-5 seconds with KaVo DRYspray.
See also: Instructions for use for the KaVo CLEANspray/KaVo DRYspray.
Preparation methods according to ISO 17664