5 Operation | 5.1 Attaching the medical device

16 / 28

5 Operation

Note

At the beginning of each workday, the water-conducting systems should be
rinsed for at least 2 minutes (without transmission handpieces being at-
tached) and if there is a risk of contamination from reflux or back suction, the
system may also need to be rinsed for 20 to 30 seconds after each patient.

5.1 Attaching the medical device

 WARNING

Detachment of the medical device during treatment.

A medical device that is not properly locked can release from the MULTIflex
coupling during treatment.

Before each use, check if the medical device is securely locked onto the
MULTIflex coupling.

NOTICE

Inaccurate coupling can destroy the high-pressure lamp or the LED of the
coupling or reduce its service life.

Make sure that the turbine is accurately coupled and firmly seated on the
coupling.

Mount the medical device accurately on the
MULTIflex (LUX) / MULTIflex LED coupling and push it backward until the
coupling audibly locks in the medical device.

Pull on the medical device to make sure that it is securely affixed to the
coupling.

5.2 Removing the medical device

Grasping the coupling, twist the medical device slightly and pull it off.

5.3 Inserting INTRA LUX head in shanks 2307 LN or
2320 LN

 CAUTION

Loosening of the medical device during treatment.

If the head is not properly locked in place, it can fall out during treatment.

Do not mount or remove the head while rotating. Before each treatment,
check that the head is firmly seated and that the clamping ring is tight.

Rotate the clamping ring in the direction of the arrow to the stop.

Insert the head to the stop. Make sure that the catches are properly seated.

Rotate the clamping ring in the opposite direction (-> close) and tighten it.

Содержание INTRAFLEX LUX 2 2307 LN

Страница 1: ...Instructions for use INTRAflex LUX 2 Programm 2307 LN 0 535 6190 2320 LN 0 535 6180 2325 LN 0 535 6210...

Страница 2: ...stributed by KaVo Dental Technologies LLC 11727 Fruehauf Drive Charlotte NC 28273 USA Phone 847 550 6800 Fax 847 550 6825 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo c...

Страница 3: ...LN or 2320 LN 16 5 4 Removing INTRA LUX head from shanks 2307 LN or 2320 LN 17 5 5 Insert the milling cutters or diamond grinders in contra angle 2325 L 17 5 6 Removing the cutting tools or diamond gr...

Страница 4: ...8 Tools and consumables 25 9 Terms and conditions of warranty 26 Instructions for use INTRAflex LUX 2 Programm 2307 LN 0 535 6190 2320 LN 0 535 6180 2325 LN 0 535 6210 Table of contents 4 28...

Страница 5: ...duling or if you have any questions please contact KaVo Repair Service KaVo Dental Technologies LLC 11727 Fruehauf Drive Charlotte NC 28273 USA Toll free Direct Customer Service 1 888 ASK KAVO 888 275...

Страница 6: ...nditions Temperature range Transportation and storage conditions Air pressure Transportation and storage conditions Humidity Protect from moisture Protect from impact HIBC Code Hazard levels The warni...

Страница 7: ...Instructions for use INTRAflex LUX 2 Programm 2307 LN 0 535 6190 2320 LN 0 535 6180 2325 LN 0 535 6210 1 User instructions NOTICE In cases which if not prevented could lead to material damage 7 28...

Страница 8: ...product and accessories appropriately before disposal If there is any injury to soft tissue do not continue treatment in the oral cavity with compressed air driven instruments 2 2 Technical condition...

Страница 9: ...ouch soft tissue with the handpiece head or instrument cover Do not use the medical device as a light probe Use an appropriate light probe for illumination of the oral cavity or site of preparation Af...

Страница 10: ...tions for use This medical device is intended for dental treatment only All other types of use or modifications of the product are not permitted and can be hazardous The medical device is intended for...

Страница 11: ...lf the patient and third parties from hazards to prevent contamination from the product 3 2 Technical Specifications 3 2 1 Reducing shank 2307 LN Identification 1 green ring Transmission ratio 2 7 1 D...

Страница 12: ...o 3 0 bar 29 to 44 psi Air consumption 50 to 60 Nl min Speed range 60 000 to 140 000 min 1 Idle torque at 3 0 bar 44 psi 0 32 Ncm Handpiece cutters or grinders can be used Attachable to all MULTIflex...

Страница 13: ...r Contaminated and moist cooling air can cause malfunctions Make sure that the supply of cooling air is dry clean and uncontaminated according to ISO 7494 2 4 1 Mount the MULTIflex LUX MULTIflex LED c...

Страница 14: ...Instructions for use INTRAflex LUX 2 Programm 2307 LN 0 535 6190 2320 LN 0 535 6180 2325 LN 0 535 6210 4 Start up and shut down 4 2 Checking the amount of water 14 28...

Страница 15: ...731 between the coupling and the medical device and check the following pressures Drive pressure drive pressure recommended return air pressure spray water pressure and spray air pressure Drive pressu...

Страница 16: ...ED of the coupling or reduce its service life Make sure that the turbine is accurately coupled and firmly seated on the coupling Mount the medical device accurately on the MULTIflex LUX MULTIflex LED...

Страница 17: ...rs with worn or damaged shafts Risk of injury tool may fall out during treatment Never use dental burs or diamond grinders with damaged or worn shafts CAUTION Danger of injury from cutters or grinders...

Страница 18: ...e 2325 LN WARNING Hazard from rotating dental bur or diamond grinder Lacerations and damage to the chuck system Do not touch the dental bur or diamond grinder when it is rotating Do not press the push...

Страница 19: ...be lubricated with a cotton ball wetted with KaVo Spray Press the O ring between your fingers to form a loop Push the O ring to the front and remove it Insert new O rings into the grooves 6 2 Cleanin...

Страница 20: ...tool Reprocess the medical device as soon as possible after treatment To minimise the risk of infection during reprocessing always wear protect ive gloves Remove the tool from the medical device Remov...

Страница 21: ...ds the following products based on compatibility of the materi als The microbiological efficacy must be ensured by the disinfectant manufac turer and proven by an expert opinion Approved disinfectants...

Страница 22: ...KaVo CLEANpac bag and place it on the corresponding care adapter Press the spray key for 1 to 2 seconds Care of clamping chuck Note Only possible for the INTRAflex contra angle KaVo recommends servici...

Страница 23: ...ether with the guide bush of the chuck to be ser viced against the tip of the service coupling Press the button marked with the chuck service symbol Note Close the chuck service mode Option 1 Place th...

Страница 24: ...lowing sterilisation processes Steriliser with triple pre vacuum at least 3 minutes at 135 C 275 F Drying time 16 min Steriliser using the gravity method at least 10 minutes at 135 C 275 F Drying tim...

Страница 25: ...shanks 2307 LN 2320 LN INTRAflex LUX contra angle handpiece 2325 LN Instrument stand 3 005 5204 X X Insert for IN TRAflex 3 005 3089 X X Cellulose pad 100 units 0 411 9862 X X Cleanpac 10 units 0 411...

Страница 26: ...ot be liable for defects and their consequences that have arisen or may arise from natural wear improper handling cleaning or maintenance non compliance with operating maintenance or connection instru...

Страница 27: ......

Страница 28: ...1 013 8706 Fk 20190214 01 en...

Отзывы: