7 Setup methods according to DIN EN ISO 17664
7.1 Preparation at the site of use
WARNING
Hazard from non-sterile products.
There is a risk of infection from contaminated medical devices.
▶ Observe suitable personal protective measures.
▶
Remove all residual cement, composite or blood without delay.
▶
Recondition the medical device as soon as possible after treatment.
▶
Remove the cutter or grinder from the medical device.
▶
The medical device must be dry when transported for reconditioning.
▶
Do not place it in a solution or similar.
Setup methods according to DIN EN ISO 17664
35