6.4.4 Cleaning: Automated internal cleaning
KaVo recommends thermodisinfectors in accordance with ISO 15883-1,
e.g.. Miele G 7781 / G 7881.
(Validation was performed with the program "VARIO-TD", the detergent
"neodisher
®
mediclean", the neutraliser "neodisher
®
Z" and rinse "neodish‐
er
®
mielclear" and only refers to the material compatibility with KaVo prod‐
ucts.)
▶
The program settings and cleansers and disinfectants that must be
used can be found in the instructions for use of the thermodisinfector.
▶
Directly after automated cleaning, treat the medical device with the
care products and systems provided by KaVo.
Preparation methods according to ISO 17664