Instructions for use EXTRAmatic 9040
4 Commissioning | 4.1 Preparation for commissioning
20 / 56
4 Commissioning
4.1 Preparation for commissioning
Note
Mains connection required.
Note
Prior to commissioning the unit, make sure that the voltage specified on the rating
plate and the enclosed mains cable are consistent with the existing mains connec‐
tion.
Set-up option
▪ Free
▪ In the table
▪ Under the table
Note
Note that the total permissible connected load is 3,000 Watt.
In combination with just one dust extraction point, no compressed air connection
and/or no switching module is required.
Close unused dust extraction openings with a stopper Mat. no. 06570492.
In combination with more than one dust extraction point, a compressed air connection
(input pressure min. 1 bar, max. 6 bar) is required.
The operating pressure for the suction path diverter is set at 0.8 bar via a pressure
reducer.
If the pressure is too high, the rubber membranes can become damaged.
2,5
2,0
1,5
1,0
0,5
0bar
4.2 Lifting and carrying the unit
▶ Before lifting, remove the front panel ①.