Instructions for use ESTETICA E70 Vision / E80 Vision
4 Operation | 4.12 Using the foot control
4.12.6 Preselect dentist
▶
Hold down the foot pedal and press the stirrup switch (with handpieces
stored).
ð
Up to six users can be programmed. This can be set in "User" in the "Set-
tings" tab.
4.12.7 Start and regulate instruments
Note
The footpedal is equipped with a middle centring function: The footpedal al-
ways returns to the middle position after being moved to the left or right in
the operating modes, excavation, preparation and prophylaxis.
CAUTION
Centring in the middle is effected for the wireless foot control by a po-
sitioning motor.
If the positioning motor breaks down, switching from or into the middle posi-
tion using the wireless foot control is no longer feasible. The type of treatment
can still be selected, but the foot pedal does not leave the middle position and
cannot switch into the middle position. The speed currently set on the wireless
foot control is always shown on the display of the treatment unit.
▶
Charge battery.
▶
If centring in the middle does not work despite the batteries being charged,
the positioning motor is defective. Have the positioning motor checked!
Note
Delays can arise after a function has been triggered if the wireless connection
of the wireless foot control is problematic.
▶
Take the handpiece (e.g., turbine, motor, PiezoLED, etc.) from the holder.
ð
The handpiece is active.
▶
Press the footpedal.
ð
The removed instrument runs at the set speed or intensity.
▶
Changing speed or intensity with the foot pedal.
ð
The left stop corresponds to the minimum speed/intensity.
ð
The right stop corresponds to the maximum speed/intensity.