Care instructions ESTETICA E50 Life
2 Setup methods according to ISO 17664 | 2.7 Manual reprocessing of the water-conducting system
NOTICE
Damage caused by treatment in a washer disinfector.
▶
Do not subject spittoon bowls made of glass to washer disinfector treat-
ment.
The spittoon bowl made of porcelain can be subjected to thermodisinfection.
2.7 Manual reprocessing of the water-conducting
system
Note
Using the "Water block, compact", a water bottle or the Water block DVGW
assembly kit with integrated germ reducing facility but no intensive germ re-
duction assembly kit, the water-conducting system must be rinsed manually.
The automatic hygiene functions are available through the intensive germ re-
duction assembly kit only.
WARNING
Risk of infection if the "Water block, compact" is used without addi-
tional safeguards.
Contamination of the treatment water and/or drinking water supply with
germs.
▶
With regard to the "Water block, compact" assembly kit, please note that
no germ reduction facility is installed in the unit, and take appropriate safe-
guards. KaVo recommends to use the "Water block DVGW with integrated
water germ reduction system" in combination with KaVo OXYGENAL 6
(
Mat. no. 0.489.3451
).
▶
If the Water bottle kit is used with the enclosed dosing attachment (
Mat.
no. 1.002.0287
), add the proper amount of KaVo OXYGENAL 6 (
Mat. no.
0.489.3451
) with each filling. For the correct amount, please refer to the
instructions of the dosing attachment for water germ reduction.
Manual reprocessing procedure:
▪ Before first start-up and after the device has not been used for a while
(weekends, holidays, vacation times, etc.)
▪ Before you start working
▪ After each patient (short rinsing programme of 20 seconds)
▶
Rinse all water outlets without handpieces for 2 minutes and actuate tum-
bler filler repeatedly.
▶
Before first start-up and after the device has not been used for a while,
briefly blow air through the air lines.
2.8 Automatic hygiene functions of the water
conducting system
Note
Intensive germ reduction kit and water block DVGW required.