Care instructions ESTETICA E50 Life
2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit
20 / 58
▶
Open the hose clamp
⑦
before startup.
Sterilisation kit for physiological saline solution
Note
The hose set sterile (10 pcs.) (
Mat. no. 1.012.4601
) is designed for single
use and must be changed after each application.
NOTICE
Damage caused by sterilisation.
▶
Do not sterilise instrument hoses and hose clips for saline hose.
Requirement
The hoses for the physiological saline solution are rinsed.
▶
Seal the motor holder, handpieces, and contra-angle handpieces (and in-
struments if applicable) in a sterilisation bag and sterilise them.
Note
Allow the sterilised items to cool to room temperature before using them
again.
2.4.8 Reprocessing KaVo ERGOcam One
Note
The camera handle cannot be sterilised or disinfected with heat.