![KaVo ELECTROmatic TM Скачать руководство пользователя страница 70](http://html1.mh-extra.com/html/kavo/electromatic-tm/electromatic-tm_instructions-for-use-manual_3703148070.webp)
Instructions for use ELECTROmatic TM and TMM/TMC
6 Operation | 6.5 ENDO Mode
70 / 90
If the set speed differs from the recommended value of this file, a crossed-out
factory symbol is shown.
See also:
2
6.5.7 Setting the ENDO torque
Note
Changes of speed or torque in ENDO mode are not saved permanently and
are available only until you exit from the view. In order to save the speed or
torque permanently, the speed or torque need to be set in the file editor.
See also:
2
6.5.3 Defining/changing ENDO file sequences, Page 63
This file has a torque of, e.g., 1.20 Ncm assigned to it.
▶
If the speed is active, activate the torque by touching the torque value.
▶
To change the torque pull the slider to the right or left.
The torque is limited to the set value.
Note
The ENDO notice signal is emitted once a level of 90% of the set torque is
reached.
If the set torque of the file differs from the recommended value of this file, a
crossed-out factory symbol is shown.
6.5.8 Selecting torque mode
There are three different torque modes in ENDO mode:
▪ Torque Control
▪ Auto-reverse
▪ Autorev/Forward
▶
Press the "Torque mode" control element to display a selection list.
Содержание ELECTROmatic TM
Страница 1: ...Instructions for use ELECTROmatic TM and TMM TMC...
Страница 89: ......
Страница 90: ...1 012 9280 bd 20200110 04 en...