3 Description of the product | 3.10 Symbols on the nameplate of the type 4882 power supply
32 / 90
Requirements
Class of protection
I
Protection class
IP 40
Ambient conditions
Permissible installation sites
Indoor use
Permissible ambient temperature
range
0
o
C to +40
o
C / 32
o
F to 104
o
F
Relative humidity
10 to 95% RH non-condensing
Pollution degree
2
Permissible up to
3,000 m / 9843 feet altitude
Atmospheric pressure limitation
700 to 1060 hPa
Transportation and storage conditions
Ambient temperature
-20
℃
to +80
℃
/ -4
o
F to 176
o
F
Relative humidity
10 to 95% RH non-condensing
Air pressure
700 hPa to 1060 hPa
3.10 Symbols on the nameplate of the type 4882 power
supply
The nameplate is located on the underside of the device.
Certification
TÜV Rheinland mark
UL mark for components for USA/Canada
CE mark
Product characteristics
Manufacturer
Type
Device type
Material number
IN-
PUT:
Input data: voltage, frequency, current
OUTPUT:
Output data: power, voltage, current
S/N: