background image

Instructions for use

DSE Compact 5195 / 5195 LUX

Содержание DSE Compact 5195

Страница 1: ...Instructions for use DSE Compact 5195 5195 LUX...

Страница 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Phone 49 0 7351 56 0 Fax 49 0 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Страница 3: ...es 5195 5195 LUX 13 3 6 Technical Specifications 13 3 7 Scope of delivery 15 4 Commissioning 16 4 1 Fitting and connecting 16 4 1 1 Connect power air and water 16 4 1 2 Installing the dentist element...

Страница 4: ...EN ISO 17664 30 6 1 Servicing 30 6 1 1 Draining and cleaning the water tank Venturi dust extraction 30 6 1 2 Filling the pressurised water system if available 31 6 2 Care 31 7 Safety check according t...

Страница 5: ...rotechnical Commission RI Repair instructions RK Retrofitting kit AS Assembly kit CK Conversion kit EP Enclosed parts EMC Electromagnetic compatibility PI Processing instructions 1 1 2 Symbols See the...

Страница 6: ...an only be asserted when they are immediately reported to KaVo in writing This notification must be accompanied by a copy of the invoice or delivery note on which the manufacturing number is clearly v...

Страница 7: ...e delivery receipt The recipient and the representative of the shipping company must sign this delivery receipt Without this evidence the recipient will not be able to assert a claim for damages again...

Страница 8: ...Instructions for use DSE Compact 5195 5195 LUX 1 User instructions 1 4 Transportation and storage 8 40 Protect from moisture Permissible stacking load Temperature range Humidity Air pressure...

Страница 9: ...s distinguishing between three hazard levels are used in this document to prevent personal and property damage The warning and safety notes in this document must be observed to prevent personal injury...

Страница 10: ...and unit Use the unit only in line with the manufacturer s instructions for use CAUTION Malfunctions due to electromagnetic fields The product meets the applicable requirements regarding electromagnet...

Страница 11: ...recycled or disposed of in strict com pliance with all applicable national regulations in a manner which is safe both for people and the environment If you have any questions regarding proper disposa...

Страница 12: ...oot control Tray Dentist element Effluent tank Water bottle 3 3 Dentist and assistant element The configuration of the dentist and assistant elements can vary according to design See also 2 4 1 4 Conn...

Страница 13: ...DSE Compact 5195 SN REF 1 009 6200 Made in China Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach Type DSE Compact 5195 LUX SN REF 1 009 6200 Input 9 15V 3 VA Made in China Type Device type DSE...

Страница 14: ...d loads Note Connect the device only to an external power supply approved according to IEC 60950 that has the appropriate voltage and power specifications Rated voltage 9 to 15 V DC Power consumption...

Страница 15: ...s depending on type of unit up to 400 000 rpm Ambient conditions Ambient temperature 5 to 40 C 41 to 104 F Relative humidity 30 to 80 Degree of soiling 2 Setup site Maximal 3 000 m above sea level Tra...

Страница 16: ...ies and electrical shock Route the connecting lines from the unit s connector and the power outlet appropriately such that there is no risk of tripping and such that the cable does not get damaged dur...

Страница 17: ...nect to the central suction unit Connection for compressed air and domestic water depending on model blue hose compressed air green hose domestic water Connect each socket to on site rapid coupling No...

Страница 18: ...ructions for Use of the handpiece Install triple function handpiece NOTICE Damage caused by non axial attachment No media coming through Depending on model axially connect handpiece hose and handpiece...

Страница 19: ...hose connector See also 2 4 1 4 Connect the handpieces Page 19 4 1 4 Connect the handpieces 3 2 1 4 Turbines or air motors Small suction hose if available Large suction hose Triple function handpiece...

Страница 20: ...g lines from the simulation unit to the customer site supply connectors Injuries and electrical shock Check the connecting lines from the simulation unit to the customer site supply connectors to ensu...

Страница 21: ...with a risk of getting pinched The swivel mechanism moves along when the height is adjusted Always position the patient simulator using both hands Press the patient positioning button on foot control...

Страница 22: ...40o backward nodding movement 2 1 Loosen the screw on the head of the dental patient simulator Move the head to the desired position After making the adjustment tighten the screw on the head again No...

Страница 23: ...ght or from right to left depending on the starting position Re attach the dentist element and the assistant element Make sure that the guide surface on the handpiece arm mates with groove of the elem...

Страница 24: ...on the dentist or assistant element 5 1 8 Switching off the DSE Compact 5195 5195 LUX and moving it to park position CAUTION Connecting lines from the simulation unit to the customer site supply conne...

Страница 25: ...the simulator is completed Fold in the patient simulator and guide the patient simulator on the up per body to the end of the park position Swivel the assistant element and the dentist element into t...

Страница 26: ...Operation 5 2 Using triple function handpieces 26 40 5 2 Using triple function handpieces Syringe DCI 3439 Cannula Water button Air button 3 2 1 4 Triple function handpiece One Cannula Water button gr...

Страница 27: ...ecrease the exiting air flow by applying more or less pressure on the air button or Press the water button and continuously increase or decrease the exiting water jet by applying more or less pressure...

Страница 28: ...ltaneously The following settings can be changed Speed Cooling level Use the foot control to set the speed and cooling level See also 2 5 4 Foot control C standard Page 28 Set the spray air and spray...

Страница 29: ...e spray mist suction device and saliva ejector Note If the saliva ejector is used the slide of the spray mist extraction device must be closed If the spray mist extraction device is used the slide of...

Страница 30: ...lectrical part of the unit may be done only by skilled staff or by staff trained in the factory During servicing work pull the mains plug out of the socket or disconnect the unit from the mains 6 1 1...

Страница 31: ...in clockwise direction and fill it Re insert the pressurised water bottle in counter clockwise direction Thoroughly rinse the water tank and re insert it Note The unit need not be vented The air press...

Страница 32: ...as well as in recurrent tests is required Caution Electrical safety checks have to be carried out by electrical engineering professionals or by individuals trained in electrical engineering and superv...

Страница 33: ...reases The suction unit is full Empty the suction unit Refill the pressurised water sys tem Patient simulator cannot be moved adjusted Air supply is interrupted Examine air connection Turbine not runn...

Страница 34: ...coupling Mat no 0 553 1550 465 LED coupling Mat no 1 007 3197 Suction cannula Mat no 0 236 2127 Step down ratio Mat no 0 236 2132 Flexible saliva ejector Mat no 0 236 2128 Dentist element DSE Compact...

Страница 35: ...Water supply integ rated installation kit Mat no 3 000 1274 EBS Suction central in stallation kit Mat no 2 000 1670 EBS Saliva ejector Venturi installation kit Mat no 2 000 2112 EBS pneumatic brake i...

Страница 36: ...odel LJ VZ Mat no 0 623 0411 Simulation model UJ TZ 14 Mat no 0 623 0531 Simulation model LJ TZ 14 Mat no 0 623 0471 Face mask Figure Description Material number Default Mat no 0 622 3882 Latex Mat no...

Страница 37: ...Instructions for use DSE Compact 5195 5195 LUX 10 Declaration of conformity 6 2 Care 10 Declaration of conformity 37 40...

Страница 38: ......

Страница 39: ......

Страница 40: ...1 010 6219 Fk 20190313 3 en...

Отзывы: