7 Reprocessing steps in accordance with ISO 17664 | 7.4 Manual Reprocessing
▶
Spray the disinfectant on a cloth, then wipe down the medical device and
allow the disinfectant to act according to the instructions of the
disinfectant manufacturer.
▶
Follow the instructions for use of the disinfectant.
▶
Do not bend the silicone legs of the tip apart during manual disinfection.
7.4.4 Manual internal disinfection
This product is not designed for manual internal disinfection.
7.4.5 Manual drying
This product is not designed for manual drying.