k990400 September 19, 2007
Calibration and Quality Assurance
10-3
Geometry Calibration
The Panel Calibration must be performed prior to running the
Geometry Calibration. It is recommended to run the Geometry
Calibration once a year or if image quality becomes degraded. The
Geometry Calibration also runs in both Portrait and Landscape
positions and takes about 12 to 15 minutes to complete.
11. Perform the Panel Calibration (procedure listed above).
12. Mount the Phantom Platform (shown below).
13. Ensure that the BB Phantom is level.
14. Center the BB Phantom on the platform using the alignment
holes on the platform and bottom of phantom.
15. Align the BB Phantom crosshair slits with the laser cross beams.
For an accurate alignment, the laser beams should penetrate
though the phantom crosshairs and appear on the receptor panel.
16. From the Calibration menu (already displayed from the Panel
Calibration) click the
Preview
button.
17. The settings screen is displayed, click
OK
.
Содержание 3D eXam
Страница 30: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 5 8...
Страница 46: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 6 16...
Страница 90: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 9 12...
Страница 99: ...k990400 September 19 2007 Calibration and Quality Assurance 10 9 17 The following preview screen is displayed...
Страница 126: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 11 10...
Страница 138: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 12 12 System Gantry Dimensions SIDE VIEW TOP VIEW FRONT VIEW...
Страница 161: ...k990400 September 19 2007 B 7...
Страница 162: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 8...
Страница 163: ...k990400 September 19 2007 B 9...
Страница 164: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 10...