
Instructions for use EXPERTmatic LUX E25 L Planmeca – 1.015.0660
4 Startup and shut down | 4.1 Checking the amount of water
12 / 28
4 Startup and shut down
WARNING
Hazard from contaminated products.
Infection hazard to the dentist and patient.
4
Prior to initial startup and after each use, reprocess the product and ac-
cessories.
WARNING
Dispose of the product in appropriate manner.
Infection hazard.
4
Reprocess the product and accessories before disposal.
Also refer to:
7 Reprocessing steps in accordance with ISO 17664, Page 19
Currently applicable packaging law
Dispose of and recycle the packaging appropriately in accordance with current
packaging law, employing waste management / recycling companies. Comply
with the comprehensive return system. KaVo has had its packaging licensed for
this purpose. Please comply with the regional public waste-disposal system.
CAUTION
Damage from soiled and moist cooling air/compressed air.
Contaminated and moist cooling air can cause malfunctions.
4
Make sure that the supply of cooling air is dry, clean and uncontaminated
according to DIN EN ISO 7494-2.
4.1 Checking the amount of water
CAUTION
Overheating of the tooth due to insufficient amount of cooling water.
Insufficient spray water can cause the medical device to overheat and damage
the pulp and tooth.
4
Adjust the water flow for the spray cooling to a minimum of 50 mL/min (3.1
inch
3
).
4
Check the spray water channels and if necessary clean the spray nozzles
with the nozzle needle (
Mat. no. 0.410.0921
).
4
Check water filter and replace, if necessary.
Содержание 1.015.0660
Страница 1: ...Instructions for use EXPERTmatic LUX E25 L Planmeca 1 015 0660...
Страница 28: ...1 015 0904 bd 20221122 01 en...