7 Reprocessing steps in accordance with ISO 17664 | 7.1 Preparations at the site of use
19 / 28
7 Reprocessing steps in accordance with ISO 17664
7.1 Preparations at the site of use
WARNING
Hazard from contaminated products.
Contaminated products are associated with an infection hazard.
4
Take suitable personal protective measures.
WARNING
Sharp dental bur in the medical device.
Injury hazard from sharp and/or pointed dental bur.
4
Remove the dental bur.
4
Reprocess the medical device right after treatment.
4
The medical device must be dry when transported to reprocessing.
4
To minimise the risk of infection during reprocessing, always wear protect-
ive gloves.
4
Remove the dental bur from the medical device.
4
Remove all residual cement, composite or blood immediately.
4
Do not place in solutions or similar substances.
7.2 Manual Reprocessing
WARNING
Sharp dental bur in the medical device.
Injury hazard from sharp and/or pointed dental bur.
4
Remove the dental bur.
CAUTION
Never reprocess this medical device in an ultrasonic cleaner.
Malfunction and material damage.
4
Reprocess in a washer disinfector or by hand only.
7.2.1 Manual external cleaning
Accessories required:
▪ Tap water 30
o
C ± 5
o
C (86
o
F ± 10
o
F)
▪ Brush, e.g. medium-hard toothbrush
4
Brush off under flowing tap water.
7.2.2 Manual internal cleaning
Validated manual internal cleaning (removal of residual protein) can be done
with KaVo CLEANspray.
4
Cover the medical device with the KaVo Cleanpac bag, and place it on the
corresponding care adapter.
4
Hold the can vertically.
4
Press the spray button three times for 2 seconds each time.
Содержание 1.015.0660