7 Processing steps in accordance with ISO 17664 | 7.7 Packaging

22 / 28

▶ Remove the bur.

▶ Service the product in the QUATTROcare PLUS.

See also:

2

Instructions for use KaVo QUATTROcare PLUS

Servicing the chuck

KaVo recommends cleaning and servicing the chuck system once a week using the
chuck servicing program integrated in the device.

See also:

2

Instructions for use KaVo QUATTROcare PLUS

Note

Handpieces must be taken off the service couplings before the chuck service can
be started and run.

▶ Close the front door and press the chuck service button for at least three seconds

until the spray canister control LED flashes three times consecutively.

ð

The device is in chuck service mode.

▶ Remove the service coupling chuck from the side hatch of the QUATTROcare

PLUS and attach it to coupling service station four, on the far right. A MULTIflex
adaptor must be mounted there.

▶ Press the handpiece together with the guide bush of the chuck to be serviced

against the tip of the service coupling.

▶ Press the button marked with the chuck service symbol.

Note

Close the chuck service mode.
Option 1: Place the dental handpieces in the QUATTROcare PLUS 2124 A, close
the front door and start the service procedure.
Option 2: After three minutes with no service procedure running, the device auto-
matically switches back to normal service mode.

See also:

2

Servicing with KaVo QUATTROcare PLUS

7.7 Packaging

Note

The sterilization bag must be large enough for the handpiece to fit without stretching
the bag.
The quality and use of the sterilization packaging must comply with applicable stan-
dards and be suitable for the sterilization procedure!

▶ Seal each medical device individually in a sterilization item package.

Содержание 1.011.6750

Страница 1: ...Instructions for use SMARTmatic S20 1 011 6750 SMARTmatic S20 K 1 011 6751 SMARTmatic S20 S 1 011 6752...

Страница 2: ...Distributed by KaVo Dental Corporation 11729 Fruehauf Drive Charlotte NC 28273 USA Phone 847 550 6800 Fax 847 550 6825 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Страница 3: ...g O rings on the motor coupling 16 6 2 2 Cleaning the spray tube 16 7 Processing steps in accordance with ISO 17664 17 7 1 Preparations at the site of use 17 7 2 Non fixing preliminary cleaning of the...

Страница 4: ...8 Tools and consumables 24 9 Terms and conditions of warranty 25 Instructions for use SMARTmatic S20 1 011 6750 SMARTmatic S20 K 1 011 6751 SMARTmatic S20 S 1 011 6752 Table of contents 4 28...

Страница 5: ...m Please direct all questions regarding the product service and maintenance to the KaVo Technical Service Toll free 1 888 ASK KAVO 888 275 5286 Email customerservice kavo com Please refer to the seria...

Страница 6: ...Air pressure Transportation and storage conditions Humidity limitations Protect from moisture Keep dry Fragile handle with care HIBC Code Hazard levels The warning and safety notes in this document mu...

Страница 7: ...he case of injury to soft tissue do not continue treatment in the oral cavity with instruments driven by compressed air To test use and remove the tool use a glove or finger guard 2 2 Technical condit...

Страница 8: ...g appropriate medical training can injure the patient the user or third parties Make sure that the user has read and comprehends the instructions for use Only employ the device if the user has the app...

Страница 9: ...ergonomically shaped The devices can be sterilized by the steam autoclave method Through the tube and the electrical motor connected to a dental unit the SMARTmatic handpieces equipped with a handpie...

Страница 10: ...cation described herein You need to comply with the following the applicable health and safety regulations the applicable accident prevention regulations these Instructions for use In accordance with...

Страница 11: ...11 6752 3 Description of the product 3 2 Technical Specifications To use the spray Attach the spray clip to the dental instrument as shown If you use the spray clip attach the hose to the spray clip a...

Страница 12: ...the product 3 3 Transportation and storage conditions 12 28 3 3 Transportation and storage conditions Do not store in a refrigerated environment Temperature 20 C to 70 C 4 F to 158 F Relative humidity...

Страница 13: ...from contaminated and moist cooling air Contaminated and moist cooling air can cause malfunctions Make sure that the supplied cooling air is dry clean and free of contamination in accordance with ISO...

Страница 14: ...ing the foot switch while attaching or detaching the medical device Property damage to the medical device Do not connect or remove the medical device while pressing the foot switch Lightly spray O rin...

Страница 15: ...ive speed of the tool or abrupt engage ment of the tool Property damage to tool shaft and chuck system reduction of the service life of tool and chuck system Do not operate the tool at a higher speed...

Страница 16: ...hile working Service the medical device When the speed drops or is uneven Service the medical device Missing O ring on the motor coupling Replace O ring 6 2 Troubleshooting 6 2 1 Replacing O rings on...

Страница 17: ...cessories required Demineralized water 30 C 2 C 86 F 3 6 F Nozzle needle Brush e g medium hard toothbrush Disposable syringe Note Before cleaning take the hose off the spray clip and the contra angle...

Страница 18: ...only 7 3 1 Manual external cleaning Accessories required Tap water 30 o C 5 o C 86 o F 10 o F Brush e g medium hard toothbrush Brush under flowing tap water 7 3 2 Automated external cleaning KaVo rec...

Страница 19: ...nds carrying out an final disinfection of the unpack aged item to guarantee the complete disinfection NOTICE Using the disinfectant bath or chlorine containing disinfectants Functional damage and prop...

Страница 20: ...ying Manual drying Clean the outside and inside with compressed air until no drops of water are visi ble Automated drying The drying procedure is usually part of the cleaning program of the washer dis...

Страница 21: ...ce with German RKI requirements KaVo recommends servicing the product as part of the processing after each use i e after each cleaning disinfection and before each sterilization Remove the bur Service...

Страница 22: ...rvice coupling chuck from the side hatch of the QUATTROcare PLUS and attach it to coupling service station four on the far right A MULTIflex adaptor must be mounted there Press the handpiece together...

Страница 23: ...e the product from the steam sterilizer immediately after the sterilization cycle The KaVo medical device has a maximum temperature resistance of up to 138 280 4 F Select a suitable procedure dependin...

Страница 24: ...dle 0 410 0931 Spray head INTRA KaVo Spray 0 411 9911 INTRA service coupling 1 009 6143 Spray clip Set 1 012 9230 Silicone hose sterilisable 0 065 5188 Material summary Mat No Adapter INTRAmatic CLEAN...

Страница 25: ...not be liable for defects and their consequences that have arisen or may arise from natural wear improper handling cleaning or maintenance non compliance with operating maintenance or connection inst...

Страница 26: ......

Страница 27: ......

Страница 28: ...1 010 0460 s 20170220 2 en US...

Отзывы: