5 Operation | 5.5 Changing default values
39 / 68
NOTE
The torque values may deviate by max. ± 10 % in the KaVo contra-angle
handpieces SURGmatic S201. Larger deviations are possible with other con-
tra-angle handpieces.
4
Press the plus and minus keys to set the desired value. The step width de-
pends on the transmission ratio and the range of values.
ð
Changed values are stored automatically.
The maximal torque value reached is displayed during the treatment in the
activities "Cut thread", "Place the dental implant" and "Set closure cap". The
value is re-set as soon as the motor starts again.
5.5.3 Setting the coolant flow
CAUTION
Coolant dosed incorrectly.
Tissue damage.
4
Please note the instructions for use of the attachment tool.
4
Set a sufficient coolant flow rate.
The coolant flow rate of the KaVo surgical units can be set to 4 levels or
switched off:
Displayed informa-
tion
Description
Off
Level 1 = 32 ml/min
Level 2 = 50 ml/min
Level 3 = 76 ml/min
Level 4 = 110 ml/min