Setup methods according to DIN EN ISO 17664
45
Sterilisation
Sterilisation in a steriliser in compliance with DIN EN 13060
CAUTION
Contact corrosion from moisture.
Damage to the product.
▶
Immediately remove the product from thesteam steriliser after the
sterilisation cycle!
The medical device has a max. temperature resistance of 138℃.
KaVo recommends for example
- STERIclave B 2200/ 2200P by KaVo
- Citomat/ K-series by Getinge
Autoclave three times with an initial vacuum for at least 4 minutes at 134°C
± 1