![KaVo 0.571.5020 Скачать руководство пользователя страница 20](http://html1.mh-extra.com/html/kavo/0-571-5020/0-571-5020_instructions-for-use-manual_658244020.webp)
Instructions for use SONICflex 2000 N - 0.571.5020
6 Preparation methods according to ISO 17664 | 6.8 Storage
20 / 26
CAUTION
Contact corrosion due to moisture.
Damage to product.
▶ Immediately remove the product from the steam steriliser after the sterilisation cy‐
cle!
Note
Remove the tips to be sterilized and dry the medical device.
The KaVo medical device has a maximum temperature resistance of up to 138 ℃
(280.4 °F).
Select a suitable procedure (depending on the available autoclave) from the following
sterilisation processes:
▪ Autoclave with three times pre-vacuum:
- at least 3 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4 °F)
▪ Autoclave using the gravity method:
- at least 10 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4 °F)
6.8 Storage
Prepared products must be stored, protected from germs (as far as possible) and
dust, in a dry, dark, cool room.
Note
Comply with the expiry date of the sterilised items.
Содержание 0.571.5020
Страница 1: ...Instructions for use SONICflex 2000 N 0 571 5020 Always be on the safe side...
Страница 4: ...6 8 Storage 20 7 Tools 21 Instructions for use SONICflex 2000 N 0 571 5020 Table of contents 4 26...
Страница 24: ...Instructions for use SONICflex 2000 N 0 571 5020 7 Tools 6 8 Storage 24 26...
Страница 25: ......
Страница 26: ...0 589 0402 kb 20100916 13 en...