![KaVo 0.535.6180 Скачать руководство пользователя страница 22](http://html.mh-extra.com/html/kavo/0-535-6180/0-535-6180_instructions-for-use-manual_658068022.webp)
Instructions for use INTRAflex LUX 2 Programm 2307 LN - 0.535.6190, 2320 LN - 0.535.6180, 2325 LN -
0.535.6210
7 Reprocessing steps in accordance with ISO 17664 | 7.4 Care products and systems - Servicing
22 / 28
CAUTION
Improper service and care.
Risk of injury.
▶
Service regularly with suitable agents.
Note
KaVo guarantees the proper function of KaVo products only if the care
products listed as accessories are used, since these were tested for proper
use on our products.
7.4.1 Servicing with KaVo Spray
KaVo recommends servicing the product as part of the reprocessing after each
use, i.e. after each cleaning, disinfection, and before each sterilisation.
▶
Remove the tool from the medical device.
▶
Cover the medical device with the KaVo CLEANpac bag, and place it on the
corresponding care adapter.
▶
Press the spray key for 1 to 2 seconds.
Care of clamping chuck
Note
Only possible for the INTRAflex contra-angle.
KaVo recommends servicing the chucking system once weekly.
▶
Remove the tool from the medical device.
▶
Position the tip of the spray nipple in the opening, and apply the spray.
▶
Press the spray key for 1 to 2 seconds.
7.4.2 Servicing with KaVo QUATTROcare PLUS
Cleaning and servicing device with expansion pressure for internal cleaning of
inorganic residues and optimum care.
KaVo recommends servicing the product as part of the reprocessing after each
use, i.e. after each cleaning, disinfection, and before each sterilisation.
▶
Remove the tool from the medical device.
▶
Service the product in the QUATTROcare PLUS.
See also:
2
Instructions for use KaVo QUATTROcare PLUS
Servicing of the clamping chuck
Note
Only possible for the INTRAflex contra-angle.
Содержание 0.535.6180