6.5.3 Disinfection: Machine disinfection - external and internal
KaVo recommends thermodisinfectors in accordance with ISO 15883-1
e.g. Miele G 7781/G 7881.
▶
The program settings and cleansers and disinfectants that need to be
used can be found in the instructions for use for the thermodisinfector.
▶
Directly after automated cleaning, treat the medical device with the
care products and systems provided by KaVo.
6.6 Drying
Manual drying
▶
Blow off the outside and inside with compressed air until no water drops
are visible.
Preparation methods according to ISO 17664