83
4.
Remove the two screws that fasten the furnace to its base. Those screws are
located just behind the orange sliding parts (see example below).
5.
Slowly flip the furnace to the back, about its hinges. Be careful: when the
furnace is tilted to the back, the door will want to spontaneously open,
because of its springs. Also, the safety shield might suddenly rotate.
Blower box
removal
In order to access the mold cooling blower or the optional solution agitation
module, it is necessary to remove the blower box with LCD screen.
IMPORTANT
: Always disconnect the instrument from the power outlet when
working with panels removed. 240 Volts inside! Risk of electrocution!
1.
Disconnect the instrument from the wall outlet.
2.
Move the platinumware holders in the furnace and open the safety shield.
Содержание K2 Prime