NanoParticle Nebulizer: Operation
Liquid NanoParticle Sizer User Guide
80
Version 1.0.1
2.
On
Experiments
screen, click
before an experiment name to edit the
experiment.
3.
When done editing, click
NEXT
→
button on
Edit Experiment
screen to
jump to
View Experiment
screen, where you can start the experiment by
clicking the
START
button. Alternatively, the
View Experiment
screen
can be directly accessed by clicking the
→
before an experiment name on
Experiments
screen.
Performing a Volume Standard Calibration for LNS Analyses
It is suggested to perform a volume standard calibration of the LNS system
on each day when an LNS measurement is scheduled. To run a calibration
with the LNS Volume Standard for the system, follow the following
instructions. Note: The volume standard calibration can be included as one
sample among other to-be-analyzed samples in the sample sequence of an
experiment.
1.
Click
button to the upper right corner of the Kanolysis
Home
screen. In the drop-down list, click
Methods
.