NanoParticle Nebulizer: Operation
Liquid NanoParticle Sizer User Guide
41
Version 1.0.1
Installing the Local Network
Kanolysis communicates with the LNS instruments using the ModbusTCP
protocol and thus requires that the computer running the software, as well as
each instrument, have a its corresponding network interface properly
configured. The instruments each use a wired Ethernet connection with a
static IPv4 configuration. Many configurations are possible, but a thorough
discussion of network administration is beyond the scope of this document.
Instead, Kanomax FMT recommends and supports the setup described in the
subsequent sections. If you would like to connect these instruments to an
existing network, please contact your network administrator for the
appropriate settings to enter. Be sure to use those same IP address values
when following the instructions in the “Configuring Kanolysis Software for
LNS Measurements” section.
To install the local network for Kanolysis communication, follow these
instructions:
1.
Plug one supplied Ethernet cable into the
Ethernet
port on the back panel
of the NanoParticle Nebulizer.
2.
Plug one supplied Ethernet cable into the
Ethernet
port on the back panel
of the Annular Flow Ion Mobility Classifier.
Ethernet cable